Associated content: 

Rosemount Icing Detector

The RICE is a magnetostrictive oscillation probe with a sensing cylinder 6.35 mm in diameter and 2.54 cm in length. Ice buildup on the sensing cylinder causes the frequency of oscillation to change, which can be related to the rate of ice accretion and hence the cloud liquid water content (LWC). When approximately 0.5 mm of ice has accumulated, a heater melts the ice, which is shed into the air stream. The heater cycle is approximately 5 s, and the cylinder normally requires an additional 5–10 s to cool down to a temperature where it can begin accreting ice again.

Point(s) of Contact: 

NCAR NOxyO3

The NCAR NOxyO3 instrument is a 4-channel chemiluminescence instrument for the measurement of NO, NO2, NOy, and O3. NOx (NO and NO2) is critical to fast chemical processes controlling radical chemistry and O3 production. Total reactive nitrogen (NOy = NO + NO2 + HNO3 + PANs + other organic nitrates + HO2NO2 + HONO + NO3 + 2*N2O5 + particulate NO3- + …) is a useful tracer for characterizing air masses since it has a tendency to be conserved during airmass aging, as NOx is oxidized to other NOy species.

NOx (NO and NO2), NOy (total reactive nitrogen), and O3 are measured using the NCAR 4-channel chemiluminescence instrument, previously flown on the NASA WB-57F and the NCAR C130. NO is measured via addition of reagent O3 to the sample flow to generate the chemiluminescent reaction producing excited NO2, which is detected by photon counting with a dry-ice cooled photomultiplier tube. NO2 is measured as NO following photolytic conversion of NO2, with a time response of about 3 sec due to the residence time in the photolysis cell. NO is measured with an identical time response due to use of a matching volume. NOy is measured via Au-catalyzed conversion of reactive nitrogen species to NO, in the presence of CO, with a time response of slightly better than 1 sec. O3 is measured using the same chemiluminescent reaction but with the addition of reagent NO to the sample flow. Time response for the ozone measurement is slightly better than 1 s.

Instrument Type: 
Measurements: 
Point(s) of Contact: 

Soluble Acidic Gases and Aerosols

As part of the measurement team on the NASA DC-8 we operate two related installations: a mist chamber/ion chromatograph (MC/IC) sampling/analysis system providing near real time results for selected species, and a bulk aerosol system that collects particulates onto filters for subsequent analysis. We use ion chromatography on aqueous extracts of the bulk aerosol samples collected on Teflon filters to quantify soluble ions (Cl-, Br-, NO3-, SO42-, C2O42-, Na+, NH4+, K+, Ca+, and Mg+). Filters are exposed on all level flight legs. Below 3 km exposure times are 5 minutes or less, increasing at higher altitudes to a maximum sample time of 15 minutes. Aerosols participate in heterogeneous chemistry, impact radiative transfer, and can be detected from space. Our measurements help to validate and extend retrievals of aerosol distributions and properties by MODIS, MISR and CALIPSO. In addition, several of the particle-associated ions are tracers of sources of gas and aerosol pollutants (e.g., SO42- from industrial emissions of SO2, enhancements of C2O42-, K+, and NH4+ indicate encounters with biomass burning plumes, Na+, and Cl- are tracers of seasalt, Mg2+ and Ca2+ are tracers of dust). Our system has two inlets, allowing collection of paired samples simultaneously.

Instrument Type: 
Measurements: 
Aircraft: 
Point(s) of Contact: 

Forward Scattering Spectrometer Probe

The FSSP is of that general class of instruments called optical particle counters (OPCs) that detect single particles and size them by measuring the intensity of light that the particle scatters when passing through a light beam. A Helium Neon laser beam is focused to a diameter of 0.2 mm at the center of an inlet that faces into the oncoming airstream. This laser beam is blocked on the opposite side of the inlet with an optical stop, a "dump spot" to prevent the beam from entering the collection optics. Particles that encounter this beam scatter light in all directions and some of that scattered in the forward direction is directed by a right angle prism though a condensing lens and onto a beam splitter. The "dump spot" on the prism and aperture of the condensing lens define a collection angle from about 4º - 12º.

The beam splitter divides the scattered light into two components, each of which impinge on a photodetector. One of these detectors, however, is optically masked to receive only scattered light when the particles pass through the laser beam displaced greater than approximately 1.5 mm either side of the center of focus. Particles that fall in that region are rejected when the signal from the masked detector exceeds that from the unmasked detector. This defines the sample volume needed to calculate particle concentrations.

Instrument Type: 
Point(s) of Contact: 

Scanning High-Resolution Interferometer Sounder

The Scanning High-resolution Interferometer Sounder (S-HIS) is a scanning interferometer which measures emitted thermal radiation at high spectral resolution between 3.3 and 18 microns The measured emitted radiance is used to obtain temperature and water vapor profiles of the Earth's atmosphere in clear-sky conditions. S-HIS produces sounding data with 2 kilometer resolution (at nadir) across a 40 kilometer ground swath from a nominal altitude of 20 kilometers onboard a NASA ER-2 or Global Hawk.

Instrument Type: 
Measurements: 
Point(s) of Contact: 

Radiometric Measurement System

Optics: We employed very simple optical arrangement for the radiometer. The diffuser-light trap arrangement provides a hemispherical field of view with incident radiation being collimated by the high reflectance walls of the exponential-logarithmic cavity. Enough collimation of the radiation is achieved with this design that narrow spectral bandpass interference filters can be used to select desired wavelength regions.

Electronics: The instrument electronics includes five major functional blocks. They are the detectors signal conditioning block, the data processing block, the system controller block, the shadow ring drive and control block, and the data storage block.

The signal detectors are silicon photodiodes operating in the photovoltaic mode and covering the spectral range from about 0.3 to 1.1µm. Their signals are converted into electrical voltages by low noise FET input operational amplifiers. Programmable gain amplifiers allow adjustments for dynamic range, and filter circuits condition the signals for analog to digital processing. Data processing units consist of an analog multiplexer circuit, a sample-and-hold circuit, and an analog to digital converter providing a 12-bit resolution output. The shadow ring is driven by a DC motor rotating at a constant speed. A motor controller is used to maintain motor speed. The system controller provides the timing necessary to perform all the system's tasks. It sets the shadow ring in motion and steps through the detector's outputs, maintaining the proper dynamic range for the analog to digital converter by selecting the proper amplifier gain. It also controls the analog to digital conversion and selectively stores data.

Instrument Type: 
Measurements: 
Point(s) of Contact: 

Microwave Temperature Profiler

The Microwave Temperature Profiler (MTP) is a passive microwave radiometer, which measures the natural thermal emission from oxygen molecules in the earth’s atmosphere for a selection of elevation angles between zenith and nadir. The current observing frequencies are 55.51, 56.65 and 58.80 GHz. The measured "brightness temperatures" versus elevation angle are converted to air temperature versus altitude using a quasi-Bayesian statistical retrieval procedure. The MTP has no ITAR restrictions, has export compliance classification number EAR99/NLR. An MTP generally consists of two assemblies: a sensor unit (SU), which receives and detects the signal, and a data unit (DU), which controls the SU and records the data. In addition, on some platforms there may be a third element, a real-time analysis computer (RAC), which analyzes the data to produce temperature profiles and other data products in real time. The SU is connected to the DU with power, control, and data cables. In addition the DU has interfaces to the aircraft navigation data bus and the RAC, if one is present. Navigation data is needed so that information such as altitude, pitch and roll are available. Aircraft altitude is needed to perform retrievals (which are altitude dependent), while pitch and roll are needed for controlling the position of a stepper motor which must drive a scanning mirror to predetermined elevation angles. Generally, the feed horn is nearly normal to the flight direction and the scanning mirror is oriented at 45-degrees with respect to receiving feed horn to allow viewing from near nadir to near zenith. At each viewing position a local oscillator (LO) is sequenced through two or more frequencies. Since a double sideband receiver is used, the LO is generally located near the "valley" between two spectral lines, so that the upper and lower sidebands are located near the spectral line peaks to ensure the maximum absorption. This is especially important at high altitudes where "transparency" corrections become important if the lines are too "thin." Because each frequency has a different effective viewing distance, the MTP is able to "see" to different distances by changing frequency. In addition, because the viewing direction is also varied and because the atmospheric opacity is temperature and pressure dependent, different effective viewing distances are also achieved through scanning in elevation . If the scanning is done so that the applicable altitudes (that is, the effective viewing distance times the sine of the elevation angle) at different frequencies and elevation angles are the same, then inter-frequency calibration can also be done, which improves the quality of the retrieved profiles. For a two-frequency radiometer with 10 elevation angles, each 15-second observing cycle produces a set of 20 brightness temperatures, which are converted by a linear retrieval algorithm to a profile of air temperature versus altitude, T(z). Finally, radiometric calibration is performed using the outside air temperature (OAT) and a heated reference target to determine the instrument gain. However, complete calibration of the system to include "window corrections" and other effects, requires tedious analysis and comparison with radiosondes near the aircraft flight path. This is probably the most important single factor contributing to reliable calibration. For stable MTPs, like that on the DC8, such calibrations appear to be reliable for many years. Such analysis is always performed before MTP data are placed on mission archive computers.

Instrument Type: 
Measurements: 
Aircraft: 
DC-8 - AFRC, ER-2 - AFRC, Global Hawk - AFRC, L-188C, M-55, Gulfstream V - NSF, WB-57 - JSC
Point(s) of Contact: 

Meteorological Measurement System

The Meteorological Measurement System (MMS) is a state-of-the-art instrument for measuring accurate, high resolution in situ airborne state parameters (pressure, temperature, turbulence index, and the 3-dimensional wind vector). These key measurements enable our understanding of atmospheric dynamics, chemistry and microphysical processes. The MMS is used to investigate atmospheric mesoscale (gravity and mountain lee waves) and microscale (turbulence) phenomena. An accurate characterization of the turbulence phenomenon is important for the understanding of dynamic processes in the atmosphere, such as the behavior of buoyant plumes within cirrus clouds, diffusions of chemical species within wake vortices generated by jet aircraft, and microphysical processes in breaking gravity waves. Accurate temperature and pressure data are needed to evaluate chemical reaction rates as well as to determine accurate mixing ratios. Accurate wind field data establish a detailed relationship with the various constituents and the measured wind also verifies numerical models used to evaluate air mass origin. Since the MMS provides quality information on atmospheric state variables, MMS data have been extensively used by many investigators to process and interpret the in situ experiments aboard the same aircraft.

Point(s) of Contact: 

Millimeter Imaging Radiometer

The Millimeter-wave Imaging Radiometer (MIR) is a cross-track-scanning radiometer that measures radiation at nine frequencies. In every scanning cycle of about 3 seconds in duration, it views two external calibration targets. MIR responds predominantly to atmospheric parameters like water vapor, clouds, and precipitation.

Instrument Type: 
Measurements: 
Point(s) of Contact: 

MODIS Airborne Simulator

The MODIS Airborne Simulator (MAS) is a multispectral scanner configured to approximate the Moderate-Resolution Imaging Spectrometer (MODIS), an instrument to be orbited on the NASA EOS-AM1 platform. MODIS is designed to measure terrestrial and atmospheric processes. The MAS was a joint project of Daedalus Enterprises, Berkeley Camera Engineering, and Ames Research Center. The MODIS Airborne Simulator records fifty spectral bands.

Instrument Type: 
Measurements: 
Aircraft: 
Point(s) of Contact: 

Pages